EIT LiQlyzer Residual Chlorine Online Meter

Choosing a Free Chlorine Analyzer

A chlorine sensor is made to detect how much chlorine is present in a solution.
When chlorine is dissolved in water, it can take on a variety of forms, including hypochlorous acid and chlorite. This is known as free chlorine, and when it is calculated as a percentage of chlorine demand, it produces residual chlorine. According to the Chlorine section, the pH of the measured solution has a significant impact on the ratio of hypochlorous acid to chlorite.

Colorimetric VS Amperometric

Amperometric
Every amperometric chlorine sensor uses this electrochemical process to measure chlorine.
HOCl + H+ + 2e– à Cl- + H2O
At the cathode, hypochlorous acid is reduced to chloride, which releases electrons. The sensor measures this electron flux.
There are two main styles: Membrane and Bare
Bare type
The three-electrode amperometric measurement method is employed in bare electrode systems. By adjusting for solution resistance, a potentiostat applies voltage between two electrodes. The measured current and chlorine concentration are inversely related.
Membrane type
The two-electrode approach is employed in a Membrane Style system. The Counter Electrodes and the Reference Electrode are united and kept apart from the solution by a membrane, otherwise it is virtually identical to the three-electrode approach. Electrolyte gradually diffuses into the solution and needs to be frequently replaced.
Colorimetric type
Chemicals applied to the sample react with chlorine in the colorimetric method to produce a colour. The amount of chlorine has a clear relation to how dark the colour is. These tools are referred to as Online DPD Instruments (DPD). A reading of the ppm chlorine is obtained by measuring the colour.
Residual Chlorine Online Meter LQCL-T40AC
LQCL-T40AC
Residual Chlorine Online Meter LQCL-T60AC
LQCL-T60AC
Residual Chlorine Sensor
Range: 0-20mg/L ; 0-100mg/L
Communication: 4-20mA RS485 ModbusRTU
Working Condition: 0-60¡ÆC
Power Supply: 240VAC
Installation: Panel/ Wall mounted/ Pipeline
Model LQCL-T40AC LQCL-T60AC
Accuracy ¡¾3%FS ¡¾1%FS
Relay control contacts 3A 250VAC, 3A 30VDC 5A 240VAC, 5A 28VDC, 120VAC
Protection IP54 IP65
Dimension 96*96*135(mm) 144*144*107(mm)
Hole Size 92.5*92.5(mm) 138*138(mm)
Other Functions - Data record &Curve display